Z-1948-2025: Brand Name: Solana Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M3

Recall record Z-1948-2025 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1948-2025
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA classification date
Jun 11, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was determined that there could be a false positive rate of up to 16.7% for the affected lots associated with the Bordetella parapertussis (BPP) results (only) of the assay.

Source facts

Action
On May 6, 2025, URGENT PRODUCT CORRECTION NOTIFICATION letters to customers. REQUIRED ACTION " Do not report Bordetella parapertussis (BPP) positive results when testing with the affected lot numbers. " Use the Solana Bordetella Complete Assay only for reporting Bordetella pertussis (BP) results and negative Bordetella parapertussis (BPP) results only when using the affected lots, as they are not affected. " Review the content of this communication with your Medical Director and retain this letter for your laboratory. " Complete the enclosed Confirmation of Receipt form no later than May 14, 2025. Contact Information We apologize for the inconvenience this will cause your laboratory. If you have further questions, please contact our Global Services Organization at 1-800-874-1517 or technicalsupport@quidelortho.com.
Initiated
2025-03-24
Root cause
Under Investigation by firm
Distribution
Please see the attached list "Consignee List_CL2025-120." Note: This file contains a specific tab, titled Shipping Summary, which contains the US and ex-US shipping summary. One (1) Distributor was affected. The other four(4) Distributors were sent the notification for information only. QuidelOrtho sent the notification to their consignees because QuidelOrtho drop-shipped this product to their 31 consignees (Included in the Medical Facility total below).
Recall number
Z-1948-2025
Recall status
Open, Classified
CDRH record ID
213840
FDA classified
2025-06-11
Recall event number
96870

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗