Z-1949-2011: Access Total T4 Reagent Kit, Part Number: 33800, Lot Numbers (s): All reagent pack lots The Access Tota

Recall record Z-1949-2011 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1949-2011
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Apr 12, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The recall was initiated because the concentration of total T4 in samples from pregnant women is erroneously low (20%) when measured using the Access Total T4 assay. Erroneously low results which are actually high (outside of the normal reference range) could mask hyperthyroidism. Failure to diagnose or treat hyperthyroid disease associated with pregnancy could result in serious, permanent harm

Source facts

Action
Beckman Coulter sent an Urgent Product Correction letter dated March 20, 2011, to all affected consignees. The letter identified the product, the problem and the action to be taken by the consignee. Consignees were issued the following instructions: (1) If you suspect pregnant patient results were evaluated at your laboratory using only Total T4 to report thyroid status, consider reviewing Total T4 results reported in the last 9 months. Consult with your Laboratory Medical Director and requesting physician to resolve any concerns you may have regarding previously reported patient results during this time period. (2) Do not use the Access Total T4 assay as the only marker for evaluating pregnant patients for thyroid disorders. (3) Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded Access Total T4 reagent kits (Part Number 33800) to another laboratory, please provide a copy of this letter to them. (4) Please complete and return the enclosed Response Form within 10 days so we are assured you have received this important communication. For questions regarding this notice, consignees were instructed to contact Beckman Coulter Customer Support Center. (1) Via our website, http://www.beckmancoulter.com/customersupport/support (2) Via phone, call 1-800-854-3633 in the United States and Canada (3) Outside the United States and Canada please contact your local Beckman Coulter
Initiated
2011-03-10
Root cause
Other
Terminated
2012-08-10
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Argentina Armenia, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Djibouti, Egypt, El Salvador, Ethiopia, France, Germany, Greece, Guatemala, Hong Kong, India, Israel, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Kuwait, Lebanon, Libyan Arab Jamahiriya, Liechtenstein, Malaysia, Mexico, Republic of Moldova, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Pakistan, Panama, Philippines, Poland, Puerto Rico, Reunion, Romania, Russian Federation, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Yemen, and Zambia
Recall number
Z-1949-2011
Recall status
Terminated
CDRH record ID
98573
FDA classified
2011-04-12
Recall event number
58169

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗