Z-1949-2026: Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Cat

Recall record Z-1949-2026 for Hologic, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1949-2026
Company in FDA record
Hologic, Inc
FDA classification date
Apr 24, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.

Source facts

Action
On March 23, 2026, Urgent Medical Device Recall letters were sent to customers. Actions to be taken: If you are receiving this notification, our records indicate that you have received a Genius" Review Station Monitor (CMP-01669). As this field correction applies to all monitors, please follow the instructions below and complete the Customer Response Form (CRF) using the link or QR code provided below. If you are a distributor or otherwise supply Genius" Review Station Monitors to downstream customers, you are responsible for promptly forwarding this notification to all affected sub-accounts and ensuring completion of the CRF for each end-user site. Do not change monitor settings " Continue to use the Genius" Review Station only with the original monitor settings configured by Hologic at installation. " Do not: " Disable Barco Application Appearance Manager (AAM) " Change Barco QAWeb calibration or compliance test settings " Adjust or otherwise modify the monitor configuration parameters outside of the original configuration by Hologic at installation. " Only Hologic, or its authorized provider may modify Genius" Review Station monitor configuration settings. If monitor settings have been altered " If you are aware that monitor display or calibration settings have been changed from the original configuration, do not use the Genius" Review Station monitor for clinical review until the monitor has been returned to the original configuration by Hologic or its authorized service provider. " Complete the Customer Response Form (CRF) so Hologic and its field service partner, IQVIA MedTech, can coordinate restoration of the approved configuration. Please promptly fill out the Customer Response Form (CRF) to acknowledge you have received this notification. Distributors should ensure that all downstream customers complete the CRF or provide the required information on their behalf.
Initiated
2026-03-23
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide and the countries of Asia Pacific, Europe, and the Middle East.
Recall number
Z-1949-2026
Recall status
Open, Classified
CDRH record ID
218900
FDA classified
2026-04-24
Recall event number
98487

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗