Z-1950-2024: HA FlexTrak II-Patient transport functionality to transport the patient from the preparation area to the

Recall record Z-1950-2024 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1950-2024
Company in FDA record
Philips North America Llc
FDA classification date
Jun 4, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Oil may leak from the HA FlexTrack Trolley onto the floor when pressing the hydraulic pedal, resulting in a potential slipping/falling hazard

Source facts

Action
Philips NA issued URGENT Medical Device Correction letter to US consigneess dated May 3, 2024 via FedEx. Philips Markets Organizations are responsible for distributing the letters outside of the US. Letter states reason for recall, health risk and action to take: A. If you see hydraulic fluid (oil) leakage: " First remove fluid from the floor using a paper towel. Then remove any remaining oil film from the floor using a cleaning agent (e.g. soap and water) and paper towel. " Contact your local Philips service representative. B. You may continue to use your HA FlexTrak Trolley in accordance with the intended use. C. Circulate this notice to all users of this device so that they are aware of the issue and associated hazard/harm until this issue has been resolved. D. Please post this letter near the associated MR system(s) for ease of reference. E. Complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the URGENT Medical Device Correction, understanding of the issue, and required actions to be taken. A Philips representative will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site to inspect the HA FlexTrack Trolley and correct if required (reference FCO78100580). If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377, 8AM-8PM EST, Monday-Friday).
Initiated
2024-05-03
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Fiji, Finland, France, French Guiana, Germany, Guadeloupe, Honduras, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Jordan, Lithuania, Martinique, Mexico, Morocco, Netherlands, Norway, Oman, Peru, Poland, Russian Fed., Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.
Recall number
Z-1950-2024
Recall status
Open, Classified
CDRH record ID
207680
FDA classified
2024-06-04
Recall event number
94603

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗