Z-1950-2026: Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number: CI-5

Recall record Z-1950-2026 for Advanced Bionics, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1950-2026
Company in FDA record
Advanced Bionics, LLC
FDA classification date
Apr 27, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by not requiring the use of a tool to replace the button/coin cell battery.

Source facts

Action
On 03/24/2026, the firm emailed customers an "URGENT: MEDICAL DEVICE RECALL M ZN-AIR BATTERY PAK (CI-5501)" Letter to inform them that, Advanced Bionics is voluntarily recalling the non-tamperproof battery door from the packaged configuration for all M Zn-Air Battery Pak (CI-5501), due to the non-tamperproof battery door does not require the use of a tool to replace the button batteries and therefore could allow unintended access to the button cell batteries within the battery compartment. Customer are instructed to: 1. Identify the Non-Tamperproof Battery Door - confirm which door is installed by performing the following check: -When inserted into the battery cartridge, the non-tamperproof door can be removed without using a tool. -The tamperproof battery door requires a tool for removal once installed. If you must use a tool to remove the door, you have the tamperproof door installed. 2. Remove and Dispose of the Non-Tamperproof Battery Door - Once identified, ensure the non-tamperproof battery door is removed from the battery cartridge. Do not reinstall or use this component. 3. Complete the Acknowledgement & Disposal Form (required) - After disposing of the non-tamperproof battery door, confirm that the component has been removed from service and disposed of appropriately by completing the Acknowledgement & Disposal Form linked in your email. For questions - contact customer service at 1-877-829-0026 or email customerservice@advancedbionics.com
Initiated
2026-03-23
Root cause
Under Investigation by firm
Distribution
Worldwide - U.S Nationwide distribution including in the state of AK, AL, AR, AZ, CA, CO, CT, CD, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Canada, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, China, Chile, Croatia, Cuba, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Kazakhstan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Trinidad and Tobago, United Arab Emirates, and United Kingdom.
Recall number
Z-1950-2026
Recall status
Open, Classified
CDRH record ID
219419
FDA classified
2026-04-27
Recall event number
98678

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗