The "leads out" connector had bent pins which would not allow connection to the Medtronic Physio Control Defibrillator cable during a product demonstration.
Source facts
Action
The firm sent a recall notice to their distributors on 12/3/2008 asking them to quarantine product and return it to HeartSync. The letter stated that customers' current inventory would be exchanged for new product. Customers were to call 1-800-828-4681 to provide the necessary information.
Initiated
2008-12-05
Root cause
Equipment maintenance
Terminated
2011-04-13
Distribution
Nationwide Distribution
Recall number
Z-1951-2011
Recall status
Terminated
CDRH record ID
97493
FDA classified
2011-04-12
Recall event number
57815