During surgical procedure the product has the potential for separation of the plug guide.
Separation of the plug guide may compromise the ability of the inserter to fully drive the implant completely into the bone, which may also compromise the ability to tension the suture properly.
Source facts
Action
On 9/30/2010, Arthrocare notified customers via letters explaining the reason for the recall and requesting the product be returned.
Initiated
2010-09-30
Root cause
Process control
Terminated
2011-08-01
Distribution
Product was released for distribution in US.
Recall number
Z-1952-2011
Recall status
Terminated
CDRH record ID
97522
FDA classified
2011-04-12
Recall event number
57830