Skip to content

Public FDA record

Z-1952-2018: Gyrus ACMI Reusable Active Cord for Bovie/ValleyLab Generators, Catalog No C650-129A (00821925008700) Th

Recall record Z-1952-2018 for Gyrus Acmi, Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1952-2018
Company in FDA record
Gyrus Acmi, Incorporated
FDA record creation date
May 29, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for the cord to spark and cause a fire.

Source facts

Action
Notification letters were distributed on April 17, 2018. The letters instruct customers to perform the following: Immediately assess any affected product you have in stock and replace the IFU with the revised IFU. Also ensure that any potential users of the devices are aware of the additional warning in the revised IFU. Future versions of the IFU will include the warning. Remove and discard all cords in use greater than one year. If you need additional copies of the IFU, call your Olympus customer service representative at 1-888-524-7266 to arrange for them to be provided. Please note on the enclosed Reply Form that you have received this information. Fax the completed Reply Form to 484-896-7128 regardless of whether you have any affected inventory at your facility. In addition, if you may have further distributed this product, please identify your customers, inform them at once of this action, provide the revised IFU and appropriately document your notification process. Your notification to your customers may be enhanced by including a copy of this letter. For further questions, please call (508) 804-2600.
Initiated
2018-04-17
Root cause
Device Design
Terminated
2023-09-18
Distribution
Worldwide Distribution - USA (nationwide) Distribution to the states of : AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, and WY., and to the countries of : Japan, Canada, Mexico, Australia, and EU.
Recall number
Z-1952-2018
Recall status
Terminated
CDRH record ID
163258
FDA record created
2018-05-29
Recall event number
79762

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗