Z-1952-2020: TandemHeart pump is assembled into kits: ProtekDuo Kit - DL3, Product: 5720-3118 - Product Usage: prov

Recall record Z-1952-2020 for Cardiac Assist, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1952-2020
Company in FDA record
Cardiac Assist, Inc
FDA record creation date
May 11, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Failure to prime due to an assembly error

Source facts

Action
LivaNova dba Tandem Life notified consignees via 1st class mail (certified return receipt) on March 06th, 2020, states reason for recall, health risk, and action to take: Check your inventory for the presence of potentially affected TandemHeart pump contained within your ordered kit as per Table 1 provided in Attachment 1; - Ensure any affected kit containing the potentially affected TandemHeart pump still in inventory gets quarantined before they can be returned. 2. Please fill and return Response Form by email to LivaNova.FSCA@livanova.comto confirm appropriate corrective actions have been executed. Please return the Response Form even if each potentially affected device listed in Attachment 1 has already been used. 3. Potentially affected product still in your inventory will be replaced once the Response Form has been returned to LivaNova. Leverage your TandemLife representativeto aid in the collection of the affected quarantined kit containing the potentially affected TandemHeart pump. Questions regarding this Medical Device Removal, please contact your TandemLife representative, or TandemLife Customer Quality at LivaNova.FSCA@livanova.com.
Initiated
2020-03-06
Root cause
Under Investigation by firm
Terminated
2020-11-20
Distribution
US Nationwide distributions.
Recall number
Z-1952-2020
Recall status
Terminated
CDRH record ID
180518
FDA record created
2020-05-11
Recall event number
85322

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗