Reason stated in the source
The firm has received complaints that the saddle is too long. The saddle is supposed to be between 7/8 and 1 1/8 inches long. The saddle in the complaints was up to 1.9 inches long.
Source facts
- Action
- Biomet Sports Medicine issued an "Urgent Medical Device Recall Notice" dated March 31, 2010 to user facilities and distributors. A second notification containing updated information was issued June 2, 2010. Consignees were instructed to discontinue use of the product and return it to the firm along with a completed Response Form. For further information, contact Biomet at 1-574-371-3755 or 1-574-371-3756, Monday through Friday, 8AM to 5PM.
- Initiated
- 2010-03-31
- Root cause
- Process control
- Terminated
- 2012-08-06
- Distribution
- Worldwide Distribution -- United States, Canada, Australia, The Netherlands, Belgium, Italy, France, New Zealand, Austria, United Kingdom, Chile, Greece, Poland, Denmark, Turkey, Portugal, South Africa and Spain.
- Recall number
- Z-1953-2010
- Recall status
- Terminated
- CDRH record ID
- 91756
- FDA classified
- 2010-07-01
- Recall event number
- 55776