Reason stated in the source

Failure in a retractor blade due to corrosion. In the area of the blade where the bearing/button meets the blade body, corrosion can form, weakening the welded joint and may cause the bearing/button to break.

Source facts

Action
The customer communication letter, dated February 26, 2018, was mailed or emailed. The letter describes the concern and the user action. A copy of the revised Instructions for Use will be provided with the letter. All consignees in the US and outside of the US who have not been previously contacted will receive notification.
Initiated
2018-02-26
Root cause
Component design/selection
Distribution
Worldwide distribution. US nationwide, South Korea, Australia, COSTA RICA, Taiwan, New Zealand, Italy, Canada, Netherlands, Norway, Sweden, Switzerland, Israel, Japan, Egypt, South Africa, United Kingdom
Recall number
Z-1953-2018
Recall status
Open, Classified
CDRH record ID
164233
FDA record created
2018-05-29
Recall event number
80024

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗