- FDA identifier
- Z-1955-2010
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA classification date
- Jul 6, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Toshiba America Medical Systems (TAMS) is initiating a field correction on the SSH-880CV Aplio Artida System due to a problem related to a wrong measurement value display in using with DICOM viewer. When an image is transferred from this system in DICOM transfer and a measurement is performed on a Doppler image in the DICOM viewer, measured values such as VTI (Velocity Time Integral) and PI/RI
Source facts
- Action
- The software will be replaced. To accomplish this, Toshiba America Medical Systems (TAMS) will notify consignees via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS for retention. For questions or concerns, please contact your Toshiba representative or the firm at (714) 730-5000.
- Initiated
- 2009-09-16
- Root cause
- Software design
- Terminated
- 2012-03-21
- Distribution
- Nationwide ((CA, GA, IA, LA, MA, MD, MN, MT, PA, PR, OH, OR, TN).
- Recall number
- Z-1955-2010
- Recall status
- Terminated
- CDRH record ID
- 91753
- FDA classified
- 2010-07-06
- Recall event number
- 55774