Z-1955-2019: MAHURKAR and Argyle acute hemodialysis catheters, Material numbers 8813793009 8813793013 8813794005 88137

Recall record Z-1955-2019 for COVIDIEN LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1955-2019
Company in FDA record
COVIDIEN LLC
FDA record creation date
Jul 9, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The priming volume values printed on the MAHURKA and Argyle acute catheters and Instructions for Use (IFU) are higher than the volumes required to fill each lumen.

Source facts

Action
On April 29, 2019, consignees were notified by letter via Federal Express and/or federal mail. The Field Safety Alert informed customers that the priming volume values printed on the MAHURKAR and Argyle acute catheters and Instructions for Use are higher than the volumes required to fill each lumen. The firm will update the IFU for the affected products to provide clarity on the printed priming volumes. In the meantime, the firm recommends the use of a non-heparinized lock solution, such as 4% sodium citrate or 1000 U/mL heparin to mitigate potential risk associated with unintended administration of additional concentrated heparin. The firm asks that customers complete and return the acknowledgement form accompanying the notification even if they do not have affected inventory. If you have experienced quality problems or adverse events, you may contact the firm at quality.assurance@covidien.com.
Initiated
2019-04-29
Root cause
Error in labeling
Terminated
2024-01-09
Distribution
Nationwide domestic distribution. Foreign distribution to Canada, Australia, New Zealand, Latin America, and Hong Kong.
Recall number
Z-1955-2019
Recall status
Terminated
CDRH record ID
172734
FDA record created
2019-07-09
Recall event number
82802

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗