Z-1955-2021: Synaptive Trackable Suction Set Standard and Malleable

Recall record Z-1955-2021 for Synaptive Medical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1955-2021
Company in FDA record
Synaptive Medical Inc
FDA record creation date
Jun 25, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to visible burrs/metal filings affixed along the suction tube inner perimeter.

Source facts

Action
On April 23, 2021 the firm sent a letter to its consignees with the following instructions: 1. Quarantine and discontinue use of all Synaptive Trackable Suction devices. 2. Contact Synaptive Recall Support at 1.866.853.8001 or by email at RecallSupport@synaptivemedical.com. Customer service will schedule an on-site inspection of each device. If this is not possible, arrangements will be made to have all impacted units returned to Synaptive. 3. Once on-site inspection or return of your stock has been arranged, please complete and return the attached Acknowledgement and Receipt Form. All forms should be submitted by email to RecallSupport@synaptivemedical.com. Note: If you are in North America and your account is supported by a Synaptive Clinical Applications Specialist (CAS), then a company representative may have already visited your site and conducted the necessary inspections. If so, we kindly ask that you still complete the Acknowledgement and Receipt Form for our records. 4. Upon inspection, all units will be graded as pass or fail. Units that pass meet normal specifications and are safe for continued use. These devices will be returned and released back into circulation. Failing (i.e. defective) units will be removed immediately and replaced by Synaptive at no cost.
Initiated
2021-04-23
Root cause
Vendor change control
Terminated
2024-01-12
Distribution
Domestic Distribution: IN, SC, MD, OR, TX ,LA, MI, NY, NV, FL, VA, CA,WI, PA and NC. International Distribution: Australia, Canada, Belgium, Netherlands, Singapore, Switzerland, Thailand, UK, Vietnam
Recall number
Z-1955-2021
Recall status
Terminated
CDRH record ID
187466
FDA record created
2021-06-25
Recall event number
87937

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗