Open original FDA record ↗Recall record
- FDA identifier
- Z-1955-2021
- Company in FDA record
- Synaptive Medical Inc
- FDA record creation date
- Jun 25, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to visible burrs/metal filings affixed along the suction tube inner perimeter.
Source facts
- Action
- On April 23, 2021 the firm sent a letter to its consignees with the following instructions: 1. Quarantine and discontinue use of all Synaptive Trackable Suction devices. 2. Contact Synaptive Recall Support at 1.866.853.8001 or by email at RecallSupport@synaptivemedical.com. Customer service will schedule an on-site inspection of each device. If this is not possible, arrangements will be made to have all impacted units returned to Synaptive. 3. Once on-site inspection or return of your stock has been arranged, please complete and return the attached Acknowledgement and Receipt Form. All forms should be submitted by email to RecallSupport@synaptivemedical.com. Note: If you are in North America and your account is supported by a Synaptive Clinical Applications Specialist (CAS), then a company representative may have already visited your site and conducted the necessary inspections. If so, we kindly ask that you still complete the Acknowledgement and Receipt Form for our records. 4. Upon inspection, all units will be graded as pass or fail. Units that pass meet normal specifications and are safe for continued use. These devices will be returned and released back into circulation. Failing (i.e. defective) units will be removed immediately and replaced by Synaptive at no cost.
- Initiated
- 2021-04-23
- Root cause
- Vendor change control
- Terminated
- 2024-01-12
- Distribution
- Domestic Distribution: IN, SC, MD, OR, TX ,LA, MI, NY, NV, FL, VA, CA,WI, PA and NC. International Distribution: Australia, Canada, Belgium, Netherlands, Singapore, Switzerland, Thailand, UK, Vietnam
- Recall number
- Z-1955-2021
- Recall status
- Terminated
- CDRH record ID
- 187466
- FDA record created
- 2021-06-25
- Recall event number
- 87937