Z-1956-2016: Access 2 Immunoassay System, Remanufactured, Catalog No. 386220 is an invitro diagnostic device used for

Recall record Z-1956-2016 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1956-2016
Company in FDA record
Beckman Coulter Inc.
FDA record creation date
Oct 11, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to the lack of updated thermal components, these instruments may be more susceptible to the internal case temperature exceeding the 18 to 36 degrees Celsius operating range, causing SYS flags. The instrument will not report results until the internal case temperature returns to the operating range. This situation may cause a delay in the reporting of patient results.

Source facts

Action
An Urgent Medical Device Recall letter dated 5/17/16 was sent to customers to inform them that Beckman Coulter is recalling the Access 2 Immunoassay System, Remanufactured instrument, because it may not have all the updated thermal components installed. The letter informs customers of the impact and actions to be taken. The letter informs the customers that their Service Representative will contact them to schedule an onsite visit to inspect their instrument and update as needed. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions are instructed to contact Customer Technical Support Center: http://www.beckmancoulter.com, 1-800-854-3633 in the United States and Canada. Customers outside the United States and Canada are instructed to contact their local Beckman Coulter representative.
Initiated
2016-05-17
Root cause
Under Investigation by firm
Terminated
2016-08-26
Distribution
US, Morocco, Venezuela
Recall number
Z-1956-2016
Recall status
Terminated
CDRH record ID
146785
FDA record created
2018-10-11
Recall event number
74212

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗