Z-1956-2018: MagNA Pure 24 System, GMMI: 07290519001 Product Usage: The MagNA Pure 24 System is a fully automated cl

Recall record Z-1956-2018 for Roche Molecular Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1956-2018
Company in FDA record
Roche Molecular Systems, Inc.
FDA record creation date
May 30, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cross-contamination of samples has been reported when running the existing protocol on the system with the rack-based purification run.

Source facts

Action
Customers were notified of the field action by letter on 4/27/18. The letter included the following required actions: " If you used the MagNA Pure 24 Pathogen 200 and/or Pathogen 1000 protocols to extract nucleic acids for an assay for a chronic infectious disease (e.g., Hepatitis C Virus), then suspected false positives should be evaluated further following local procedures and retested as necessary. " Complete the attached fax form and fax it to 1-844-861-6232 or email it to roche6612@stericycle.com. " File this Urgent Medical Device Correction (UMDC) for future reference.
Initiated
2018-05-02
Root cause
Software in the Use Environment
Terminated
2019-04-30
Distribution
Worldwide Distribution - US Nationwide in the states of: CA, GA, HI, IN, MD, ME, MI, MN, NC, NY, PA, VI, and WI. The products were distributed to the following foreign countries: Austria, Belgium, Brazil, Chile, China, Columbia, Costa Rica, Czech Republic, Denmark, France, Germany, Guatemala, Hong Kong, Hungary, India, Israel, Italy, Japan, Mexico, Netherlands, Norway, Poland, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, UAE, United Kingdom, Vietnam.
Recall number
Z-1956-2018
Recall status
Terminated
CDRH record ID
163850
FDA record created
2018-05-30
Recall event number
79899

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗