Open original FDA record ↗Recall record
- FDA identifier
- Z-1956-2020
- Company in FDA record
- Merit Medical Systems, Inc.
- FDA record creation date
- May 11, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
labeling error: Due to a manufacturing issue, product package was incorrectly labeled with the incorrect sheath size and the incorrect sheath assembly was included in the product packaging. The use of the incorrect kit components may result in vasoconstriction, moderate hemorrhage, and/or embolism.
Source facts
- Action
- A verbal (telephone) notification attempt was made to the customer on September 12th and 13th. These attempts resulted in a voicemail informing the customer of a voluntary product recall and requesting a call back. A hospital employee was reached by telephone on September 17, 2019 and notified of the product recall. The notification included product identification information, instruction to immediately quarantine any devices and to discontinue use and to return product to Merit per the instructions contained in the Customer Response Form (CRF). The CRF and details of the recall were emailed to the hospital per customer request.. Merit Medical Systems, Inc. is conducting a recall of the K21-00031 Custom Prelude IDeal Kit, lot number H1583914, due to the kit including a Prelude EASE sheath assembly rather than the labeled Prelude IDeal sheath assembly. Additionally, the kit has a labeling error in which the unit and carton labeling indicate a 6 French color (green) rather than the correct 4 French color (red). Merit Medical has not received any complaints from customers or any reports of patient harm or injury as a result of this issue. Return the affected products to Merit via Federal Express 2nd Day Air (Acct# 112952756), include the assigned RMA number. Send to Field Assurance Department Merit Medical Systems, Inc. 1600 West Merit Parkway South Jordan, UT 84095 Attn: RMA# (found in contact information table) For further information contact Merit Customer service at (801) 208-4381 or by email at response@merit.com.
- Initiated
- 2019-09-12
- Root cause
- Process control
- Terminated
- 2022-05-27
- Distribution
- US Nationwide distribution in the state of MI.
- Recall number
- Z-1956-2020
- Recall status
- Terminated
- CDRH record ID
- 180571
- FDA record created
- 2020-05-11
- Recall event number
- 85342