Z-1956-2021: Trilogy 100, Trilogy 200, Garbin Plus, Aeris, LifeVent Continuous Ventilators.

Recall record Z-1956-2021 for Philips Respironics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1956-2021
Company in FDA record
Philips Respironics, Inc.
FDA classification date
Jul 15, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.

Source facts

Action
Urgent Medical Device Recall letters were sent to customers starting 6/14/21. 1. Do not stop or alter your prescribed therapy until you have talked to your physician. Your physician may evaluate if the benefit outweighs the risks. 2. If your physician determines that you must continue using this device, use an inline bacterial filter. Your Instructions for Use has guidance on installation. 3. Register your device on www.philips.com/src-update or call 1-877-907-7508. An updated URGENT: Medical Device Recall notification letter was issued on 11/16/21. 1. Do not stop or change ventilator use until you have talked to your health care provider. If patients requiring a ventilator for life-sustaining therapy have no alternate, or therapy disruption is unacceptable, the HCP may determine if the benefit of continued use of these ventilator devices outweighs the potential risks identified in the recall letter. 2. Talk to your HCP about using an inline bacterial filter to help filter out foam particles. Instructions for Use has guidance on installation. At this time, the FDA does not have evidence of the safety and effectiveness of a filter for mitigating the foam risks, and evaluation is ongoing. Filters will not reduce exposure to chemicals from the PE-PUR foam. Filters may increase air flow resistance and affect ventilator performance. Monitor for accumulation of foam debris in the filter or resistance-related problems in the breathing circuit after filter placement. 3. Report any problems with a device through the FDA MedWatch Voluntary Reporting Form. 4. Register your device on the recall website www.philips.com/src-update with your device Serial Number, or call 1-877-907-7508. Devices should be serviced only by qualified technicians. There are no user serviceable parts. Attempts to remove the sound abatement foam may render the device permanently inoperative. Devices damaged due to foam removal attempts will not be able to be remediated.
Initiated
2021-06-14
Root cause
Device Design
Distribution
Worldwide distribution including US Nationwide, Argentina, Australia, Brazil, Canada, China, France, Germany, Italy, Japan, Korea, and Latin America.
Recall number
Z-1956-2021
Recall status
Open, Classified
CDRH record ID
187816
FDA classified
2021-07-15
Recall event number
88071

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗