Skip to content

Public FDA record

Z-1957-2016: The Red Head 2 Reusable Fat Harvesting Canister, Product # RH-2, Non-Sterile. Intended for use in aspir

Recall record Z-1957-2016 for Miami Fat Supply, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1957-2016
Company in FDA record
Miami Fat Supply, Inc
FDA record creation date
Jun 14, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During an FDA inspection it was found that the products are marketed without a cleared 510k, nor is approved for multiple use.

Source facts

Action
Customers were sent an Urgent: Medical Device Recall/Correction letter, dated 04/19/2016, on 04/21/2016 that served as a notification of the Red Head 2 voluntary recall. Customers were informed that there were previous complaints of compromised product integrity and performance following sterilization. Lids are subject to fracture and fail after steam sterilization. Customers were advised to destroy used product and to ship unused product back to the manufacturer. Customers were to return an acknowledgement form to confirm their notification of the voluntary recall.
Initiated
2016-04-21
Root cause
No Marketing Application
Terminated
2016-12-14
Distribution
Worldwide Distribution -- US, including the states of AL, AZ, CA, CO, FL, GA, IL, MD, MI, MO, NM, NY, OH, TX, UT, and VA; and, the country of Dominican Republic.
Recall number
Z-1957-2016
Recall status
Terminated
CDRH record ID
146527
FDA record created
2016-06-14
Recall event number
74156

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗