- Recall classification
- Class II
Reason stated in the source
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector
Source facts
- Action
- Siemens Healthcare issued A Customer Safety Advisory Notice dated 4/20/20 via AX030/20/S. Letter states reason for recall, health risk and action to take: We strongly recommend that you do not disconnect the main unit from the monitor trolley when the system displays the system error message Err 16305 / 80: Confirm this error and repeat your last action until the problem has been corrected by a service engineer. Contact our service organization at 1-800-888-7436. Siemens is currently developing a solution to eliminate the root cause of this problem.
- Initiated
- 2020-04-20
- Root cause
- Process design
- Terminated
- 2021-05-10
- Distribution
- US Nationwide distributions.
- Recall number
- Z-1957-2020
- Recall status
- Terminated
- CDRH record ID
- 181228
- FDA classified
- 2020-05-04
- Recall event number
- 85553