Open original FDA record ↗Recall record
- FDA identifier
- Z-1957-2025
- Company in FDA record
- Microbiologics Inc
- FDA classification date
- Jun 11, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Affected products were packaged with the incorrect KWIK STIK assembly resulting in the product containing the wrong microorganism. 0621K was packaged 01065K swabs; 01065K was packaged with 0621K swabs. This may lead to delayed diagnosis.
Source facts
- Action
- On May 27, 2025, the firm began notifying customers via Urgent Medical Device Recall letters. Customers were informed of the affected products containing the wrong microorganisms. Customers were instructed to review lab procedures to understand how this information affects their usage, and then to use or discard the product depending on the lab procedures and how the issue affects usage. Distributors of the affected product were instructed to contact end users who have received the affected items.
- Initiated
- 2025-05-27
- Root cause
- Mixed-up of materials/components
- Distribution
- US Nationwide distribution in the states of California and Pennsylvania and the country of New Zealand.
- Recall number
- Z-1957-2025
- Recall status
- Open, Classified
- CDRH record ID
- 214140
- FDA classified
- 2025-06-11
- Recall event number
- 96974