Skip to content

Public FDA record

Z-1958-2011: Dental chairs distributed under several brand names such as DCI Equipment; Marus MaxStar; Marus NuStar; a

Recall record Z-1958-2011 for Dental Equipment LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1958-2011
Company in FDA record
Dental Equipment LLC
FDA classification date
Apr 12, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Link arm assembly that mounts the rear mounted accessories to the dental chair may potentially break and fall off the dental chair.

Source facts

Action
On 03/08/11, Dental Equipment LLC. started sending out the Medical Device Recall notification letter via USPS 1st class mail to US consignees and Federal Express mail to foreign consignees. The firm also sent a recall notification letter to consignees who only purchased dental accessories such as Rear Mounted Cuspidor with Coverset, Arm Articulating Rear Mounted Cuspidor - Marus Gray, Maxstar Orbit Cuspidor utility W/ Porcelain bowl, and zinc plated bolts. The letter identified the affected dental chair as Marus brand with the following model numbers: DC1490, DC1690, DC1700 and DC1702. The firm's corrective action including the replacement of the affected link arm assembly with a new link arm assembly. The firm's field representative will contact each consignee for an appointment to replace the affected link arm assembly. Consignees can contact the Dental Equipment LLC at 503-537-3617 for any questions.
Initiated
2011-03-01
Root cause
Process control
Terminated
2011-09-21
Distribution
Worldwide distribution to USA, Australia, Canada, Ireland, South Africa, and Vietnam.
Recall number
Z-1958-2011
Recall status
Terminated
CDRH record ID
98425
FDA classified
2011-04-12
Recall event number
58122

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗