Open original FDA record ↗Recall record
- FDA identifier
- Z-1958-2011
- Company in FDA record
- Dental Equipment LLC
- FDA classification date
- Apr 12, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Link arm assembly that mounts the rear mounted accessories to the dental chair may potentially break and fall off the dental chair.
Source facts
- Action
- On 03/08/11, Dental Equipment LLC. started sending out the Medical Device Recall notification letter via USPS 1st class mail to US consignees and Federal Express mail to foreign consignees. The firm also sent a recall notification letter to consignees who only purchased dental accessories such as Rear Mounted Cuspidor with Coverset, Arm Articulating Rear Mounted Cuspidor - Marus Gray, Maxstar Orbit Cuspidor utility W/ Porcelain bowl, and zinc plated bolts. The letter identified the affected dental chair as Marus brand with the following model numbers: DC1490, DC1690, DC1700 and DC1702. The firm's corrective action including the replacement of the affected link arm assembly with a new link arm assembly. The firm's field representative will contact each consignee for an appointment to replace the affected link arm assembly. Consignees can contact the Dental Equipment LLC at 503-537-3617 for any questions.
- Initiated
- 2011-03-01
- Root cause
- Process control
- Terminated
- 2011-09-21
- Distribution
- Worldwide distribution to USA, Australia, Canada, Ireland, South Africa, and Vietnam.
- Recall number
- Z-1958-2011
- Recall status
- Terminated
- CDRH record ID
- 98425
- FDA classified
- 2011-04-12
- Recall event number
- 58122