Z-1958-2014: Simplexa" Flu A/B & RSV Direct, Model Number: MOL2650. 510(k) K120413 Simplexa" Flu A/B & RSV Dire

Recall record Z-1958-2014 for Focus Diagnostics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1958-2014
Company in FDA record
Focus Diagnostics Inc
FDA classification date
Jul 1, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Focus Diagnostics is providing an urgent safety notice for a correction to the labeling for Simplexa" Flu A/B & RSV Direct (MOL2650). Focus Diagnostics received some customer complaints of Simplexa Flu A/B & RSV Direct assays with sporadic false signals, which may result in a higher false results rate due to a potential software spectral matrix and Direct Amplification Disc storage.

Source facts

Action
Focus Diagnostics sent an Urgent Safety Notice Correction letter dated May 15, 2014, to all affected customers. The notification, informed customers of the product, the problem, and the action to be taken by the customer. To decrease the likelihood of obtaining any false results caused by either manually entering a spectral matrix or storing the DAD improperly after each use, please follow the recommended warnings and precautions provided in the revised and attached Simplexa Flu A/B & RSV Direct (PI.MOL2650 Rev. D) package insert. 1. The spectral matrix must be installed in each 3M Integrated Cycler and should not be changed unless an updated QR code for the instrument is provided by Focus Diagnostics. The spectral matrix is unique to each 3M Integrated Cycler. The spectral matrix was provided with the 3M Integrated Cycler instrument on the cover of the 3M Integrated Cycler Hardware Manual. If the matrix label will not scan or cannot be found, contact Focus Diagnostics. The contact information is on the last page of this document. 2. Not installing or changing the spectral matrix can result in false results. 3. After each use store DAD discs flat with the numbered foil side up. Customers were instructed to acknowledge that they have received this notification by signing the enclosed acknowledgement form and email the form to DxTS@focusdx.com or fax back to Focus Diagnostics Technical Services at 562-240-6526 within 10 business days. The Notice of Correction and Acknowledgement Form was sent to customers on 05/15/2014 via email and USPS mail. Additional customers were notified on 05/29/2014 via email and USPS mail. For questions regarding this recall call 714-220-1900.
Initiated
2014-05-15
Root cause
Device Design
Terminated
2014-12-31
Distribution
Nationwide Distribution
Recall number
Z-1958-2014
Recall status
Terminated
CDRH record ID
127387
FDA classified
2014-07-01
Recall event number
68347

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗