Z-1958-2019: Radrex-i X-Ray System (DRAD-3000E) TFP-4336W (Wireless FPD) Software version: v5.00ER005 Product Usage:

Recall record Z-1958-2019 for Canon Medical System, USA, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1958-2019
Company in FDA record
Canon Medical System, USA, INC.
FDA record creation date
Jul 11, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

software malfunction; It was found when a user performs radiography using the wireless flat panel detector (FPD), a message window displays on the monitor stating image transmission was not completed and there was no image available. It also showed the "OK" button to reacquire image data from the FPD, and the "Cancel" button to cancel the acquisition . When the user selects the "OK" button, the same message window appears. This prompted the user to repeat the same operation several times and finally select the "Cancel" button to quit the reacquistion mode.

Source facts

Action
On October 16, 2019, the firm sent an urgent medical device correction letter via USPS, return receipt to all customers who were distributed the affected device. The notification letter ask customers to do the following: 1. Should the problem occur before the modified software is installed on your system, please reduce the number of other devices which use the same wireless channel as this system as much as possible to avoid heavy wireless traffic. 2. Please share this information with all users and reviewing radiologist as well as clinical engineering and the biomedical group at your facility 3. Please complete and return the customer reply form via email to RAffairs@us.medical.canon or by fax to regulatory affairs at 877-349-3054. 4. If you have any questions regarding the letter, please contact Terry Schultz, Regulatory Affairs Manager at (800) 421-1968 or contact your local Canon Representative at (800) 521-1968.
Initiated
2018-10-16
Root cause
Software design
Terminated
2020-07-20
Distribution
US Nationwide Distribution - PA, UT, FL, NY, AR, LA, WV, NJ, OH, MI, WI, TN, TX, VT, GA, NC, IL
Recall number
Z-1958-2019
Recall status
Terminated
CDRH record ID
173970
FDA record created
2019-07-11
Recall event number
83193

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗