Open original FDA record ↗Recall record
- FDA identifier
- Z-1958-2019
- Company in FDA record
- Canon Medical System, USA, INC.
- FDA record creation date
- Jul 11, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
software malfunction; It was found when a user performs radiography using the wireless flat panel detector (FPD), a message window displays on the monitor stating image transmission was not completed and there was no image available. It also showed the "OK" button to reacquire image data from the FPD, and the "Cancel" button to cancel the acquisition . When the user selects the "OK" button, the same message window appears. This prompted the user to repeat the same operation several times and finally select the "Cancel" button to quit the reacquistion mode.
Source facts
- Action
- On October 16, 2019, the firm sent an urgent medical device correction letter via USPS, return receipt to all customers who were distributed the affected device. The notification letter ask customers to do the following: 1. Should the problem occur before the modified software is installed on your system, please reduce the number of other devices which use the same wireless channel as this system as much as possible to avoid heavy wireless traffic. 2. Please share this information with all users and reviewing radiologist as well as clinical engineering and the biomedical group at your facility 3. Please complete and return the customer reply form via email to RAffairs@us.medical.canon or by fax to regulatory affairs at 877-349-3054. 4. If you have any questions regarding the letter, please contact Terry Schultz, Regulatory Affairs Manager at (800) 421-1968 or contact your local Canon Representative at (800) 521-1968.
- Initiated
- 2018-10-16
- Root cause
- Software design
- Terminated
- 2020-07-20
- Distribution
- US Nationwide Distribution - PA, UT, FL, NY, AR, LA, WV, NJ, OH, MI, WI, TN, TX, VT, GA, NC, IL
- Recall number
- Z-1958-2019
- Recall status
- Terminated
- CDRH record ID
- 173970
- FDA record created
- 2019-07-11
- Recall event number
- 83193