Open original FDA record ↗Recall record
- FDA identifier
- Z-1959-2012
- Company in FDA record
- Cardiovascular Systems, Inc.
- FDA classification date
- Jul 9, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
CSI has initiated a recall on specific lot numbers of Stealth 360-degree Orbital Atherectomy devices which have the potential for an insufficient crown solder bond that may result in the crown becoming detached from the driveshaft.
Source facts
- Action
- Phone calls to affected sites began June 15, 2012 to quarantine the product. These same sites were also notified with an Urgent Medical Device Recall letter, dated June 18, 2012, via FedEx and e-mail (where available) on June 18, 2012. The letter notified each customer to remove from use and return the recalled product. The customer was instructed to complete a Customer Acknowledgement Form and return it with the device(s) being returned in the Returned Goods Authorization kit provided. Customers with questions can contact customer service at 1-877-274-0901. For any questions regarding this recall call 651-259-2805.
- Initiated
- 2012-06-18
- Root cause
- Employee error
- Terminated
- 2012-07-19
- Distribution
- Nationwide distribution including CA, FL, IL, MI, MN, PA, and TX.
- Recall number
- Z-1959-2012
- Recall status
- Terminated
- CDRH record ID
- 110453
- FDA classified
- 2012-07-09
- Recall event number
- 62406