Open original FDA record ↗Recall record
- FDA identifier
- Z-1959-2015
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Jul 2, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When performing a 2D Stress Echo study, some following keystrokes result in a potential loss of data, where some selected clips may not be saved as part of the study.
Source facts
- Action
- Siemens sent an Customer Safety Advisory Notification on May 29,2015, to all affected customers. The letter identified the product the problems and the action needed to be taken by the customer. Customer Safety Advisory notifications inform users of the potential issue with the ACUSON SC2000 systems with software versions between VA16! and VA30A and the Stress Echo feature. Letters describe the steps that result in the potential loss of data and suggest that the following steps be taken to remedy the situation: 1. Select Stress Echo Selection 2. If applicable relabel the clips for all Stages and Views 3. Do not return to Stress Echo Selection to select additional relabeled clips. The issue will be resolved in an upcoming software release for customers who did not receive the original software update. Until the necessary repairs have been completed, all personnel in the Organization who need to be aware of the issue should be notified. For further questions please call (650) 694-5398.
- Initiated
- 2015-05-29
- Root cause
- Software design
- Terminated
- 2017-03-09
- Distribution
- Worldwide Distribution - US Distribution and to the countries of : Albania, Angola, Armenia, Australia, Austria, Bangladesh, Belarus, Bolivia, Bosnia Herzog, Brazil, Brunei, Bulgaria, Canada, Colombia, Czech Republic, Ecuador, Egypt, Finland, France, Germany, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Lithuania, Malaysia, Maldives, Mauritius, Mexico, Netherlands, Nigeria, Oman, P.R. China, Philippines, Poland, Portugal, Qatar, Rep. of Yemen, Republic Korea, Romania, Russian Fed, San Marino, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Suriname, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., United Kingdom and Venezuela
- Recall number
- Z-1959-2015
- Recall status
- Terminated
- CDRH record ID
- 137900
- FDA classified
- 2015-07-02
- Recall event number
- 71469