Mislabeled outer carton:
Monopolar Probe is listed as 200 x 1.5mm when it should be listed as 280 x 1.5mm
Source facts
Action
Spine Wave sent an Urgent: Medical Device Recall letter dated June 11, 2015 to affected customers. The letter identified the affect product, problem and actions to be taken. Customers were instructed to acknowledge receipt
of notification by returning the enclosed Recall Notice Acknowledgement Form in the stamped envelope provided.
For questions, contact Spine Wave directly at 203-944-9494
and ask for the Recall Coordinator or by email at rsmith@spinewave.com.
Initiated
2015-05-04
Root cause
Packaging change control
Terminated
2017-01-24
Distribution
US Nationwide Distribution in the state of CA, FL, GA, LA, MA, MD, MI, MO, NC, NH, OH, OR, PA, TX, and VA.
Recall number
Z-1960-2015
Recall status
Terminated
CDRH record ID
137960
FDA classified
2015-07-02
Recall event number
71497