Open original FDA record ↗Recall record
- FDA identifier
- Z-1960-2020
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA record creation date
- May 11, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a manufacturing variation in the arm that can result in a persistent recoverable fault error (error 23087) and loss of arm functionality prior or during surgery.
Source facts
- Action
- On 11/21/2018, the firm sent an "Urgent Medical Device Correction" Notification to its customers informing them that the firm had become aware of specific surgical systems were shipped with a manufacturing variation in the arm, which may result in a persistent recoverable fault (error 23087) with could occur prior or during a surgical procedure. The firm emphasized that as indicated in the User Manual, if this fault is persistent, ten the affected arm may be disabled. The firm is instructing that: 1. Inform surgeons and patient side assistants using the surgical system to read and understand the contents of this letter. 2. Additionally, forward this letter to your Risk Manager, OR Director, Purchasing, Biomedical Engineering staff, and other members of your medical staff who perform procedures. 3. Place a copy of this letter with their User Manual until the corrective action has been completed. 4. Inform affected personnel when corrective action has been completed. 5. Please log into the Recalling Firm's Online Community Field Action resource to read or complete any requested actions related to this issue, at this link: https://www.davincisurgerycommunity.com/ 6. In the case where the online resource cannot be used, complete the attached Acknowledgement Form and return it via fax. 7. Please retain a copy of the letter and the acknowledgement form for their files. Actions being taken by the Recalling Firm are: 1. To correct the issue described in this Field Safety Notice, an Recalling Firm representative will schedule a site visit to replace the affected arm. 2. A copy of the Field Safety Notice will be provided to customers with affected surgical systems. 3. A Recalling Firm's representatives will be available by phone to answer any questions related to this Field Safety Notice. Contact Information for any questions: If you need further information or support concerning this Medical Device Notification , please contact your Clinical Sales Repres
- Initiated
- 2018-11-21
- Root cause
- Process control
- Terminated
- 2022-05-31
- Distribution
- U.S.: CA, MD, NY, VA, IL, MI, DE, TX, FL, NJ, MA, KY and OH. O.U.S.: Australia, France, Germany and Japan.
- Recall number
- Z-1960-2020
- Recall status
- Terminated
- CDRH record ID
- 180598
- FDA record created
- 2020-05-11
- Recall event number
- 84023