Open original FDA record ↗Recall record
- FDA identifier
- Z-1961-2009
- Company in FDA record
- B. Braun Interventional Systems
- FDA classification date
- Sep 3, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
B. Braun Interventional Systems was notified by B. Braun Medical France, the manufacturer of the Celsite Access port that they had received 11 reports of polyurethane (type P) catheter disconnections. B. Braun Medical France has decided to recall all Celsite ports supplied with the affected catheter lot. To date, B. Braun Interventional Systems (US) has not received any reports of catheter dis
Source facts
- Action
- B. Braun Interventional Systems, Inc. issued an "Urgent: Medical Device Recall" notice dated June 22, 2009. The letter identified the problem and product. The letter reminded the consignee of the general instruction for use of access ports and instructed them to contact patients who had the Celsite port system implanted. For further questions, contact B. Braun Interventional Systems, Inc. at 1-800-523-9676 extension 4818.
- Initiated
- 2009-06-22
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-06-05
- Distribution
- Nationwide Distribution -- NY.
- Recall number
- Z-1961-2009
- Recall status
- Terminated
- CDRH record ID
- 83457
- FDA classified
- 2009-09-03
- Recall event number
- 52611