Open original FDA record ↗Recall record
- FDA identifier
- Z-1961-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Apr 12, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Overheating. A software bug may cause the transducer to overheat when used with certain transducer models. The transducer can become noticeably hot within a few minutes, which may cause patient discomfort or burns.
Source facts
- Action
- The firm, SIEMENS, sent a "Customer Safety Advisory Notification" to its customers on 3/21/2009. The notification described the product, problem and actions to be taken. The customers were instructed to disconnect any transducers that may be in an active transducer port and chose another imaging transducer prior to use of the Aux CW transducer. If they do not have another transducer to select other than those listed in the notification, they should close the study in progress and start a new study before using the Aux CW transducer-this procedure should be followed for each new patient and /or exam type. In addition, the customers were instructed to pass this notice on to all those within their organization who need to be aware of this issue until the corrective action is completed. SIEMENS will release a software update that will correct the issue. The customers local service engineer will install the updates on their system as soon as it becomes available. Should you have any questions regarding this notification, you may contact your local service support person and/or call (650) 969-9112.
- Initiated
- 2009-03-21
- Root cause
- Software design
- Terminated
- 2012-03-09
- Distribution
- Worldwide distribution.
- Recall number
- Z-1961-2011
- Recall status
- Terminated
- CDRH record ID
- 84291
- FDA classified
- 2011-04-12
- Recall event number
- 52928