Z-1961-2012: Stryker brand Arthroscope and Hardware Set Tray (sterilization tray); Model/Part Number: 233-032-116; Pro

Recall record Z-1961-2012 for Stryker Endoscopy. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1961-2012
Company in FDA record
Stryker Endoscopy
FDA classification date
Jul 9, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the "Instructions for Use"; due to the firm's evaluation that identified original validation was not done at worst case conditions.

Source facts

Action
Stryker Endoscopy sent a Urgent Device Correction letter dated June 19, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to: 1. Discard any old Instructions for Use for your Arthroscope and Hardware Set Tray. 2. Download the new IFU from the URL link: http://stryker.com/arthrotray . Complete and sign the enclosed acknowledgement of receipt form and fax to (408) 754-8378 or scan in and email to trays@stryker.com.* It is important to send the self addressed confirmation form back to Stryker. Stryker is committed to delivering the highest quality products to its customers. We sincerely apologize for any inconvenience this action may cause. If you have any questions regarding this letter please contact us at 1-800-624-4422 or via email at trays@stryker.com. .
Initiated
2012-06-19
Root cause
Process control
Terminated
2012-10-23
Distribution
Worldwide Distribution--USA (nationwide)
Recall number
Z-1961-2012
Recall status
Terminated
CDRH record ID
110301
FDA classified
2012-07-09
Recall event number
62346

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗