Open original FDA record ↗Recall record
- FDA identifier
- Z-1961-2014
- Company in FDA record
- Integra LifeSciences Corp.
- FDA classification date
- Jul 1, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Integra Lifesciences is recalling NewPort System MIS Rod, 2 because it may be up to 30 degrees offset from specification requirements causing misalignment with the inserter. This could potentially result in the lordosis of the rod not aligning with the handle when the inserter is used in accordance with the current NewPort MIS System Surgical Technique.
Source facts
- Action
- A customer notification letter was sent on 6/9/14 to their customers and distributors who purchased the Integra NewPort System MIS Rod, 2 to inform them there may be a potential that the hex on some NewPort System MIS Rod, 2 are up to 30 degrees offset from specification requirements causing misalignment with the inserter. Customers are instructed to complete and return the customer response form and return by email or fax as indicated on the form. Customers with questions are instructed to contact Customer Service at (866) 942-8698.
- Initiated
- 2014-06-09
- Root cause
- Under Investigation by firm
- Terminated
- 2014-11-06
- Distribution
- Worldwide Distribution. US including states of CA, FL, IL, MD, MO, TN, TX, and countries of Australia, Czech Republic and Spain.
- Recall number
- Z-1961-2014
- Recall status
- Terminated
- CDRH record ID
- 128087
- FDA classified
- 2014-07-01
- Recall event number
- 68498