Public FDA record

Z-1961-2020: da Vinci X Surgical System

Recall record Z-1961-2020 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1961-2020
Company in FDA record
Intuitive Surgical, Inc.
FDA record creation date
May 11, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a manufacturing variation in the arm that can result in a persistent recoverable fault error (error 23087) and loss of arm functionality prior or during surgery.

Source facts

Action
On 11/21/2018, the firm sent an "Urgent Medical Device Correction" Notification to its customers informing them that the firm had become aware of specific surgical systems were shipped with a manufacturing variation in the arm, which may result in a persistent recoverable fault (error 23087) with could occur prior or during a surgical procedure. The firm emphasized that as indicated in the User Manual, if this fault is persistent, ten the affected arm may be disabled. The firm is instructing that: 1. Inform surgeons and patient side assistants using the surgical system to read and understand the contents of this letter. 2. Additionally, forward this letter to your Risk Manager, OR Director, Purchasing, Biomedical Engineering staff, and other members of your medical staff who perform procedures. 3. Place a copy of this letter with their User Manual until the corrective action has been completed. 4. Inform affected personnel when corrective action has been completed. 5. Please log into the Recalling Firm's Online Community Field Action resource to read or complete any requested actions related to this issue, at this link: https://www.davincisurgerycommunity.com/ 6. In the case where the online resource cannot be used, complete the attached Acknowledgement Form and return it via fax. 7. Please retain a copy of the letter and the acknowledgement form for their files. Actions being taken by the Recalling Firm are: 1. To correct the issue described in this Field Safety Notice, an Recalling Firm representative will schedule a site visit to replace the affected arm. 2. A copy of the Field Safety Notice will be provided to customers with affected surgical systems. 3. A Recalling Firm's representatives will be available by phone to answer any questions related to this Field Safety Notice. Contact Information for any questions: If you need further information or support concerning this Medical Device Notification , please contact your Clinical Sales Repres
Initiated
2018-11-21
Root cause
Process control
Terminated
2022-05-31
Distribution
U.S.: CA, MD, NY, VA, IL, MI, DE, TX, FL, NJ, MA, KY and OH. O.U.S.: Australia, France, Germany and Japan.
Recall number
Z-1961-2020
Recall status
Terminated
CDRH record ID
180599
FDA record created
2020-05-11
Recall event number
84023

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗