The devices were mislabeled with the incorrect product size resulting in 16Fr devices being labeled as 14Fr devices.
Source facts
Action
The recalling firm issued Field Safety Notices dated 11/19/2019 with FSN Ref and FSCA Ref numbers of #0000000-11152019-001-R titled "Mislabelled 16Fr Introducer Sheaths Labelled as 14FR (3 lots)" issued via email and overnight mail. Return of the devices was requested.
Initiated
2019-11-18
Root cause
Labeling Change Control
Terminated
2020-05-13
Distribution
All distribution is foreign to the following countries: Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Japan, Norway, Spain, Sweden, Switzerland, and the UK.
Recall number
Z-1962-2020
Recall status
Terminated
CDRH record ID
180716
FDA record created
2020-05-11
Recall event number
85379