If the universal ingredient (UI) in an active configuration is changed using the Configuration Editor, a flush of the outlet pump tube will not be initiated by the software. It could result in the original UI remaining in the tube which could be delivered into the next bag.
Source facts
Action
An Urgent Device Correction communication via 1st class mail was mailed on July 14, 2014, informing customers of a potential issue if the universal ingredient in an active configuration is changed using the Configuration Editor rather than following the normally prescribed and supported methods as outlined in the Operator Manual.
Initiated
2014-06-10
Root cause
Software design
Terminated
2017-02-22
Distribution
US Nationwide
Recall number
Z-1963-2014
Recall status
Terminated
CDRH record ID
128106
FDA classified
2014-07-01