Arrow is recalling the product due to a labeling inconsistency. One portion of the lidstock states the kit contains a 3-Lumen catheter while the rest of the lidstock correctly identifies the catheter as 2-Lumen.
Source facts
Action
An "Urgent Medical Device Recall Notification" letter dated July 31, 2013 was sent to direct accounts requesting immediate discontinuation of use and quarantine of all affected product on hand. Negative responses were also requested and the immediate return of product to Arrow was authorized. Questions were directed to Customer Service at 800 343-2935.
Initiated
2013-07-31
Root cause
Labeling Change Control
Terminated
2015-04-28
Distribution
AL, PA, CT, NC, VA, KS, RI, AZ, NJ, and IL.
Recall number
Z-1965-2013
Recall status
Terminated
CDRH record ID
120336
FDA classified
2013-08-14
Recall event number
65829