Z-1966-2025: Allon 2001. Part Number: 200-00263

Recall record Z-1966-2025 for Belmont Instrument LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1966-2025
Company in FDA record
Belmont Instrument LLC
FDA classification date
Jun 12, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms are disabled, or 2. The Water In and Water Out and Thermostat sensors are disconnected from the Human Sensor Board AND the temperature alarms are disabled, the device will proceed to heat the water above the hardware fault limit without notifying the user of the faulty condition.

Source facts

Action
Belmont Medical Technologies notified consignees via letter on about 05/13/2025. Consignees were instructed to ensure users of the devices are made aware of the correction, identify affected units on hand and check their serial numbers, ensure all devices have the Temp Sensor Alarms always set to ON to ensure that the user is alerted to errors and conditions requiring actions and attach the provided Step by Step Guide (P/N DDT-063-027) to each device. Consignees were also instructed to complete and return the response form and notify customers if the units were further distributed.
Initiated
2025-05-13
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of CA, DC, LA, MA, NM, PA, TX, VA, WA and the countries of Belgium, Bolivia (Plurinational State of), Brazil, Canada, Colombia, Israel, Italy, Portugal, Romania, Sweden, Taiwan (Province of China), Thailand, United Kingdom.
Recall number
Z-1966-2025
Recall status
Open, Classified
CDRH record ID
213802
FDA classified
2025-06-12
Recall event number
96854

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗