Reason stated in the source
GE Healthcare has recently become aware of an incident where the wrong part of the patient body was scanned during a CT standalone exam with the Discovery NM/CT 570c
Source facts
- Action
- GE Healthcare will remedy the issue or bring the product into compliance with each applicable Federal standard. GE Healthcare has provided field modification instructions (FMI) to update software for the Discovery NM/CT570c hybrid system. The software update will correct the incorrect positioning of the table not prescribed by the operator. The FMI includes: A software upgrade with foreign language support to be installed in the affected models listed above, Labels (FMI label and packing label), Software upgrade workflow procedure with functional verification for all modes of operation, and Certification Report, Postponement of Medical Device Correction, Rejection of Medical Device Correction, or Refusal and/or Location Unknown Documentation. For further questions please call (262) 513-4122.
- Initiated
- 2013-04-10
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2014-01-21
- Distribution
- US Distribution in the state of CT.