Z-1967-2023: bellavista 1000e Ventilator, Catalogue Number 301.100.130; Continuous Use Ventilator

Recall record Z-1967-2023 for Vyaire Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1967-2023
Company in FDA record
Vyaire Medical
FDA classification date
Jun 15, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang). When this occurs, if the user continues to interact with the touchscreen, a message will appear that states DeviceSoftware.Application is not responding or a decommission screen will appear on the user interface. The ventilator will issue both an audible and visual alarm. Ventilation continues without interruption with the settings applied prior to the APP Hang error. 2. The potential for a use error has been identified if the operator applies the proposed settings without confirming the settings are suitable for the patient.

Source facts

Action
The firm notified its consignees by email on 04/27/2023. The notice explained the issue, the risk to health and requested the following: ACTIONS TO BE TAKEN BY CONSUMERS/END USERS 1. Confirm receipt and thoroughly review the contents of the Customer Notification package. 2. If affected devices are transferred to another location or organization, forward the complete Customer Notification package to the respective users. 3. To reduce the change that the AppHang error will occur, Vyaire recommends that when the )2 suction maneuver is active, the end user should not attempt to activate any controls or screens until: The O2 suction maneuver has timed or The O2 suction maneuver has been aborted by selecting the O2 suction button again to turn it off. Vyaire further recommends the following: User re-configures the screen to remove the circuit monitoring tile, if applicable. User does not connect anything to either of the two USB ports on the left side of the ventilator during patient ventilation. For more detailed instructions, please refer to Appendix A. Following these steps will help to reduce the likelihood of the AppHang occurring. 4. When transitioning between ventilation modes, always check and confirm the proposed settings. The proposed settings are based on the current level of support provided in the active mode at the time of the mode change. Special attention should be given when transitioning back from a spontaneous breathing mode such as Pressure Support Ventilation (PSV).
Initiated
2023-05-05
Root cause
Nonconforming Material/Component
Distribution
US Nationwide Distribution
Recall number
Z-1967-2023
Recall status
Completed
CDRH record ID
199976
FDA classified
2023-06-15
Recall event number
92243

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗