A software issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy.
Source facts
Action
Between June 2023 and November 2023 software updates to resolve the issue were performed by TurnCare on affected units. The software update corrected an issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy. Successful update of the software was verified on each affected controller.
Initiated
2023-11-01
Root cause
Software design
Distribution
US Nationwide distribution in the states of Texas, Illinois, Pennsylvania, Florida, Connecticut
Recall number
Z-1967-2025
Recall status
Open, Classified
CDRH record ID
213849
FDA classified
2025-06-12
Recall event number
96873