Open original FDA record ↗Recall record
- FDA identifier
- Z-1968-2023
- Company in FDA record
- Galt Medical Corporation
- FDA classification date
- Jun 15, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Sheath introducer assemblies could potentially, due to improper storage conditions, have a luer connection portion of the dilator hub that cracks or dislodges during use, which renders the dilator component unusable, and can require the use of a replacement device.
Source facts
- Action
- On 5/5/23, recall notices were sent to customers who were asked to do the following: 1) Identify and segregate the affected lot(s) that are in your possession as well as those in the possession of your end-users. 2) Complete and return the field correction reply form to quality@galtmedical.com 2) Return affected product to the recalling firm. Customers with questions can email: dderrick@galtmedical.com
- Initiated
- 2023-05-08
- Root cause
- Process design
- Distribution
- Worldwide - US Nationwide distribution in the states of PA, UT, FL, PA, OH, GA, MA, NV, CA, NY, NJ and the countries of UK, AU, TR, FR, CH, NO.
- Recall number
- Z-1968-2023
- Recall status
- Open, Classified
- CDRH record ID
- 200245
- FDA classified
- 2023-06-15
- Recall event number
- 92322