A software update corrected an issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy.
Source facts
Action
Between November 2023 and January 2024 software updates to resolve the issue were performed by TurnCare on affected units. Successful update of the software was verified on each affected controller.
Initiated
2023-08-10
Root cause
Software design
Distribution
US Nationwide distribution in the states of Texas, Illinois, Pennsylvania, Florida, Connecticut.
Recall number
Z-1968-2025
Recall status
Open, Classified
CDRH record ID
213850
FDA classified
2025-06-12
Recall event number
96874