Z-1970-2011: LANTIS oncology Information System, an accessory to the Linear Accelerator, as part of the COHERENCE RT T

Recall record Z-1970-2011 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1970-2011
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Apr 13, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

When editing the note type of a previously entered note, then reverting to the original type, data may be lost.

Source facts

Action
Siemens sent an Urgent Medical Device Correction letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to configure their system to lock all new notes automatically. To automatically lock new notes, open eChart Navigator, right mouse click in the notes section, select "configuration" and select "Automatically Lock New Notes." If the customer chooses not to configure their system to automatically lock new notes, follow at least one of the suggestions listed below: If the customer must edit a note type, use the append feature and document that they are making an edit to the note. Lock all new notes that are entered via eChart Notes. To lock a note manually, after creating the note, right mouse click and select "Lock," Access and enter patient notes via eChart Notes until a LANTIS upgrade has been performed at your site. When entering patient notes via eChart Navigator, make sure that you are entering the notes in the correct section.\ When entering patient notes via eChart Notes, make sure that the correct note type has been selected. Customers are to keep the Customer Safety Advisory Notice together with their LANTIS 8.30 User Manual where it should remain until the measures defined above are performed. For questions call 1-888-826-9702.
Initiated
2009-08-21
Root cause
Software design
Terminated
2011-04-13
Distribution
Worldwide Distribution - USA including DE, WI, TX, TN, NJ, PA, NE, MA, and AL and the countries of South Africa, Australia, Ireland, Germany, Brazil, South Korea, India, and Japan.
Recall number
Z-1970-2011
Recall status
Terminated
CDRH record ID
84912
FDA classified
2011-04-13
Recall event number
53195

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗