Open original FDA record ↗Recall record
- FDA identifier
- Z-1970-2023
- Company in FDA record
- LumiraDx
- FDA classification date
- Jun 15, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Fails Post-Sterilization Sterile Swabs Cantilever (Bend) Testing and may be more susceptible to breakage. If the swab breaks in the nasal cavity of a patient may cause injury or medical intervention to remove part of the swab
Source facts
- Action
- LumiraDx issued on 25 April 2023 via email electronic Cover Letter and Urgent Product Recall Notice with an attestation form to document product destruction at their facility. Letter states reason for recall, health risk and action to take: Immediately stop using the affected lot of swabs " Remove from your stock any remaining affected kits of swabs. " Destroy all stock of affected swabs " Indicate the number of packs that were destroyed, and sign the attestation form at the end of this notice verifying destruction of the affected swabs in your inventory. " Report any issues or adverse events with using the affected swabs to LumiraDx immediately or to the FDA via the MedWatch program. " Please forward this notice to additional testing sites if these products were further distributed within your system. " LumiraDx will provide replacement product after receipt of the attestation form. Please contact CustomerServices.US@LumiraDx.com to arrange overnight shipment. Please complete and return the enclosed product destruction and attestation form to LumiraDx as soon as possible. If you have any questions, contact LumiraDx Technical Services at Technicalservices.US@lumiradx.com or 1-888-586-4721 option 1
- Initiated
- 2023-04-25
- Root cause
- Process control
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1970-2023
- Recall status
- Open, Classified
- CDRH record ID
- 200329
- FDA classified
- 2023-06-15
- Recall event number
- 92334