When using a Laser-Indirect-Ophthalmoscope (LIO) as the delivery device, a scattered laser beam inside the laser cavity system may unintentionally exit through the Slit Lamp output port (SDS) and get emitted from the Objective Lens when the laser is fired.
Source facts
Action
Ellex Safety Bulletin is the customer notification letter. Ellex will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. For questions contact your local Ellex-Lumibird representative or email us on eservice@ellex.com.
Initiated
2023-03-16
Root cause
Radiation Control for Health and Safety Act
Distribution
US nationwide/ Worldwide Distribution
Recall number
Z-1971-2023
Recall status
Open, Classified
CDRH record ID
201305
FDA classified
2023-06-27
Recall event number
92532
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.