Due to an improper fixture revision, device may be missing a screw at the jaw assembly, resulting in jaw components falling off during use.
Source facts
Action
The firm submitted an MDR but determined that there was no risk to health. E-mails were sent on June 26, 2008, to consignees, describing the product, the problem, and action to be taken by consignees. The consignees were ask to provide a list of all unused ThermaSeal devices and their locations, to return the products to Starion from the lots noted and to provide a copy of the attached letter to their sales force and customers if they choose.
If you have any questions, please call (408) 522-5200 ext. 326 or (408) 350-9934.
Initiated
2008-06-26
Root cause
Process control
Terminated
2010-10-27
Distribution
Worldwide distribution.
Recall number
Z-1972-2010
Recall status
Terminated
CDRH record ID
91719
FDA classified
2010-07-09
Recall event number
55750