Z-1972-2023: Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Super

Recall record Z-1972-2023 for Boston Scientific Neuromodulation Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1972-2023
Company in FDA record
Boston Scientific Neuromodulation Corporation
FDA classification date
Jun 15, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.

Source facts

Action
On 05/12/23, correction notices were mailed to physicians/surgeons, hospitals, healthcare professionals who were advised the following: IFU advises against forced deployment, as it could result in device breakage. Correction notice advises to review of the IFUs/Surgical Technique Manual. Avoid application of excessive force, use only two or three fingers when performing final tightening during device deployment). Inspect Driver instrument prior to/following use; if Driver tip break is noted: Retrieve metal fragment(s), consider performing imaging. Missing/bent teeth prevents Driver engagement with implant and requires exchange for a new/replacement Driver instrument to complete successful deployment. If metal fragment(s) are not successfully removed: MRI scans are NOT advised, Append the patient s medical record with a copy of this letter to maintain awareness of metal fragment(s) remaining in situ for the remaining service life of the device. Consider alternative imaging for patients with metal fragment(s) remaining in situ, e.g., X-ray/ultrasound/CT scan. Distributors asked to notify all customers that have been shipped/sold affected product. Complete/return acknowledgement form to BSCFieldActionCenter@bsci.com or Fax to: Field Action Center (1-866-213-1806). Driver tip breakage can be identified via an audible click, a loss of tactile resistance during the implant procedure, and/or visual detection following removal of the Driver instrument from the inserter tool and subsequent inspection. Adverse events or quality concerns should be reported to BSN.ComplaintCallCenter@bsci.com. If you have any question, contact Senior Management Quality: at 651-581-0761 or email: rebecca.kinkead@bsci.com Firm is updating IFU to advise against MRI scanning for patients with unretrieved metal fragments. A separate letter was sent to health care professions providing them with a patient letter. Patient letter states firm recommends patients discuss this letter with t
Initiated
2023-05-12
Root cause
Labeling design
Distribution
US Nationwide Distribution: IN, NY, IL, PA, FL, SC, TX, MA, VA, AL, WI, OK, MO, MI, OH, WV, NC, UT, NH, CA, NJ, TN, NV, AZ, NE, IA, CT, ME, LA, KY, MS, MD, DC, GA, CO, KS, ID, MN, AR, HI, NM, WA, SD, OR, DE, WY, AK, and ND
Recall number
Z-1972-2023
Recall status
Open, Classified
CDRH record ID
199835
FDA classified
2023-06-15
Recall event number
92181

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗