- Action
- On 05/12/23, correction notices were mailed to physicians/surgeons, hospitals, healthcare professionals who were advised the following:
IFU advises against forced deployment, as it could result in device breakage.
Correction notice advises to review of the IFUs/Surgical Technique Manual.
Avoid application of excessive force, use only two or three fingers when performing final tightening during device deployment).
Inspect Driver instrument prior to/following use; if Driver tip break is noted: Retrieve metal fragment(s), consider performing imaging. Missing/bent teeth prevents Driver engagement with implant and requires exchange for a new/replacement Driver instrument to complete successful deployment.
If metal fragment(s) are not successfully removed: MRI scans are NOT advised,
Append the patient s medical record with a copy of this letter to maintain awareness of metal fragment(s) remaining in situ for the remaining service life of the device.
Consider alternative imaging for patients with metal fragment(s) remaining in situ, e.g., X-ray/ultrasound/CT scan.
Distributors asked to notify all customers that have been shipped/sold affected product.
Complete/return acknowledgement form to BSCFieldActionCenter@bsci.com or Fax to: Field Action Center (1-866-213-1806).
Driver tip breakage can be identified via an audible click, a loss of tactile resistance during the implant procedure, and/or visual detection following removal of the Driver instrument from the inserter tool and subsequent inspection.
Adverse events or quality concerns should be reported to BSN.ComplaintCallCenter@bsci.com. If you have any question, contact Senior Management Quality: at 651-581-0761 or email: rebecca.kinkead@bsci.com
Firm is updating IFU to advise against MRI scanning for patients with unretrieved metal fragments.
A separate letter was sent to health care professions providing them with a patient letter.
Patient letter states firm recommends patients discuss this letter with t
- Initiated
- 2023-05-12
- Root cause
- Labeling design
- Distribution
- US Nationwide Distribution: IN, NY, IL, PA, FL, SC, TX, MA, VA, AL, WI, OK, MO, MI, OH, WV, NC, UT, NH, CA, NJ, TN, NV, AZ, NE, IA, CT, ME, LA, KY, MS, MD, DC, GA, CO, KS, ID, MN, AR, HI, NM, WA, SD, OR, DE, WY, AK, and ND
- Recall number
- Z-1972-2023
- Recall status
- Open, Classified