Reason stated in the source

There is a potential for the driver to bind up when attempting to advance the intervertebral body fixation screw.

Source facts

Action
Urgent Medical Device Recall letters dated 6/13/19 were emailed to customers.
Initiated
2019-06-13
Root cause
Process control
Terminated
2020-05-11
Distribution
The products were distributed to the following US states: AZ, MD, MO, and FL.
Recall number
Z-1973-2019
Recall status
Terminated
CDRH record ID
173905
FDA record created
2019-07-12
Recall event number
83166

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗