McKesson Provider Technologies - Medical Imaging Group
FDA classification date
Jul 9, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
McKesson has recently discovered that certain Horizon Cardiology Hemo systems were running specific Schiller Argus-Pro PB 1000 firmware versions that might cause a delay in the display of physiological parameters on the Horizon Cariology Hemo Real Time Monitor Screen.
Source facts
Action
The firm, McKesson, sent an "Advisory Notice" letter dated April 15, 2010, to all customers. The letter describes the problem, products affected, and actions to be taken by customers. The customers were instructed not to use the Emergency Replacement Hemo kit until they receive a replacement Schiller Argus Pro PB-1000 and they have contacted McKesson Support department at 1-866-777-0202.
If you have any questions regarding this notice, please call our Customer Support department at 1-0866-777-0202 and ask to speak to your support Manager.
Initiated
2010-04-15
Root cause
Software change control
Terminated
2010-07-21
Distribution
Nationwide distribution: CA, CO, MD, NH, NT, NY, TN, and TX .
Recall number
Z-1974-2010
Recall status
Terminated
CDRH record ID
91258
FDA classified
2010-07-09
Recall event number