Skip to content

Public FDA record

Z-1974-2010: Horizon Cardiology Hemo Component Schiller Argus Pro PB-1000 Monitor unit. Argus Pro PB 1000 firmware ver

Recall record Z-1974-2010 for McKesson Provider Technologies - Medical Imaging Group. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1974-2010
Company in FDA record
McKesson Provider Technologies - Medical Imaging Group
FDA classification date
Jul 9, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

McKesson has recently discovered that certain Horizon Cardiology Hemo systems were running specific Schiller Argus-Pro PB 1000 firmware versions that might cause a delay in the display of physiological parameters on the Horizon Cariology Hemo Real Time Monitor Screen.

Source facts

Action
The firm, McKesson, sent an "Advisory Notice" letter dated April 15, 2010, to all customers. The letter describes the problem, products affected, and actions to be taken by customers. The customers were instructed not to use the Emergency Replacement Hemo kit until they receive a replacement Schiller Argus Pro PB-1000 and they have contacted McKesson Support department at 1-866-777-0202. If you have any questions regarding this notice, please call our Customer Support department at 1-0866-777-0202 and ask to speak to your support Manager.
Initiated
2010-04-15
Root cause
Software change control
Terminated
2010-07-21
Distribution
Nationwide distribution: CA, CO, MD, NH, NT, NY, TN, and TX .
Recall number
Z-1974-2010
Recall status
Terminated
CDRH record ID
91258
FDA classified
2010-07-09
Recall event number
55598

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗