Z-1974-2023: Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS (Troponin I), 97022HS (CK-MB and Troponin I), 97000HSE

Recall record Z-1974-2023 for QUIDEL CARDIOVASCULAR INC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1974-2023
Company in FDA record
QUIDEL CARDIOVASCULAR INC
FDA classification date
Jul 7, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Raw material issue may cause cardiac panel products to have an average of 30% troponin signal reduction across measurement range 0.05 ng/mL to 30 ng/mL, QC testing won't detect anomaly, so an alternate method is recommended, if unavailable, use recommendations to mitigate potential patient impact, which could be missed/delayed myocardial infarction diagnosis.

Source facts

Action
On 5/25/2023, correction notices were sent to customers and distributors who were advised the following: 1) If you have an alternate method, please discard all unused material. 2) If you do not have an alternate method, please follow these steps, as applicable, to minimize patient risk. a) Flag all negative results reported to clinicians as possibly inaccurate until lots of unaffected product are used. b) Use results from an alternate clinical laboratory analyzer when troponin results are below or close to the cutoff and myocardial infarction is suspected. c) Perform serial sampling. Keep patients until at least 3 negative troponin values have been obtained. d) Use all Triage troponin results in conjunction with the patient s risk factors, clinical presentation, EKG, and other imaging. e) Consider recommendations by the ACC, ESC guidelines and the Fourth Universal Definition of Myocardial Infarction for monitoring a patient for a rise or fall pattern of troponin. 3) If you are experiencing issues with Proficiency testing, contact your local Technical Solutions Center. 4) Complete and return the confirmation of receipt form. 5) forward this notification if the product was distributed outside of your facility. If you have further questions, please contact: - For North America, Canada, Asia-Pacific, and Latin America, please call 858.552.1100 - For Europe, Middle East, and Africa, please call +353 (91) 412 474 - For China, please call 0400 920 9366 or +86 021 3217 8300 - EU authorized representative: MDSS GmbH, Tel.: +49-511-62628630, vigilance@mdss.com On 7/12/23, Update/Reminder Notices were sent informing customers to: Discontinue use of all affected devices. Use an unaffected lot number, obtain unaffected product, use an alternate method or send patients to an alternate testing site utilizing an alternate method. If an alternate method is not available, see above.
Initiated
2023-05-25
Root cause
Nonconforming Material/Component
Distribution
US: TX, NC, CA, LA, SC, OR, HI, OK, AZ, AL, KS, WA, FL, VA, IA, WV, MT, UT, CO MS, NM, MA, NJ, MO AK, IL, MN, WI, IN, GA, NE, MD. OUS: GBR, ITA, FRA, DEU, MCO, EGY, LCA, CHE, ASM, VIR, IRL
Recall number
Z-1974-2023
Recall status
Open, Classified
CDRH record ID
200921
FDA classified
2023-07-07
Recall event number
92447

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗